Heart Rhythm
Volume 4, Issue 1 , Pages 1-4, January 2007

Potential interference of small neodymium magnets with cardiac pacemakers and implantable cardioverter-defibrillators

  • Thomas Wolber, MD

      Affiliations

    • Cardiovascular Center, Cardiology, University Hospital, Zurich
    • Corresponding Author InformationAddress reprint requests and correspondence: Thomas Wolber, M.D., Cardiovascular Center, Cardiology, University Hospital Zurich, Raemistr. 100, CH-8091 Zurich, Switzerland.
  • ,
  • Salome Ryf, PhD

      Affiliations

    • Institute for Biomedical Engineering, University and ETH, Zurich
    • Federal Office of Public Health, Berne, Switzerland.
  • ,
  • Christian Binggeli, MD

      Affiliations

    • Cardiovascular Center, Cardiology, University Hospital, Zurich
  • ,
  • Johannes Holzmeister, MD

      Affiliations

    • Cardiovascular Center, Cardiology, University Hospital, Zurich
  • ,
  • Corinna Brunckhorst, MD

      Affiliations

    • Cardiovascular Center, Cardiology, University Hospital, Zurich
  • ,
  • Roger Luechinger, PhD

      Affiliations

    • Institute for Biomedical Engineering, University and ETH, Zurich
  • ,
  • Firat Duru, MD

      Affiliations

    • Cardiovascular Center, Cardiology, University Hospital, Zurich

Received 31 August 2006; accepted 8 September 2006. published online 18 September 2006.

Background

Magnetic fields may interfere with the function of cardiac pacemakers and implantable cardioverter-defibrillators (ICDs). Neodymium-iron-boron (NdFeB) magnets, which are small in size but produce strong magnetic fields, have become widely available in recent years. Therefore, NdFeB magnets may be associated with an emerging risk of device interference.

Objective

We conducted a clinical study to evaluate the potential of small NdFeB magnets to interfere with cardiac pacemakers and ICDs.

Methods

The effect of four NdFeB magnets (two spherical magnets 8 and 10 mm in diameter, a necklace made of 45 spherical magnets, and a magnetic name tag) was tested in forty-one ambulatory patients with a pacemaker and 29 patients with an ICD. The maximum distance at which the magnetic switch of a device was influenced was observed.

Results

Magnetic interference was observed in all patients. The maximum distance resulting in device interference was 3 cm. No significant differences were found with respect to device manufacturer and device types.

Conclusion

Small NdFeB magnets may cause interference with cardiac pacemakers and ICDs. Patients should be cautioned about the interference risk associated with NdFeB magnets during daily life.

Keywords: Magnet, Pacemaker, Implantable cardioverter-defibrillator, Electromagnetic interference, Neodymium

To access this article, please choose from the options below

Login to an existing account or Register a new account.

  • Purchase this article for 31.50 USD (You must login/register to purchase this article)

    Online access for 24 hours. The PDF version can be downloaded as your permanent record.

  • Subscribe to this title

    Get unlimited online access to this article and all other articles in this title 24/7 for one year.

  • Claim access now

    For current subscribers with Society Membership or Account Number.

  • Visit SciVerse ScienceDirect to see if you have access via your institution.
 

PII: S1547-5271(06)01962-X

doi:10.1016/j.hrthm.2006.09.012

Heart Rhythm
Volume 4, Issue 1 , Pages 1-4, January 2007