Heart Rhythm
Volume 5, Issue 7 , Pages 957-964, July 2008

Biplane three-dimensional augmented fluoroscopy as single navigation tool for ablation of atrial fibrillation: Accuracy and clinical value

  • Joris Ector, MD

      Affiliations

    • Department of Cardiology, University Hospital Gasthuisberg, University of Leuven, Leuven, Belgium
    • Corresponding Author InformationAddress reprint requests and correspondence: Joris Ector, M.D., Cardiology, University Hospital Gasthuisberg, Herestraat 49, B-3000 Leuven, Belgium, Europe.
  • ,
  • Stijn De Buck, MSc, PhD

      Affiliations

    • Department of Electrical Engineering, University Hospital Gasthuisberg, University of Leuven, Leuven, Belgium
  • ,
  • Wim Huybrechts, MD

      Affiliations

    • Department of Cardiology, University Hospital Gasthuisberg, University of Leuven, Leuven, Belgium
  • ,
  • Dieter Nuyens, MD, PhD

      Affiliations

    • Department of Cardiology, University Hospital Gasthuisberg, University of Leuven, Leuven, Belgium
  • ,
  • Steven Dymarkowski, MD, PhD

      Affiliations

    • Department of Radiology, University Hospital Gasthuisberg, University of Leuven, Leuven, Belgium.
  • ,
  • Jan Bogaert, MD, PhD

      Affiliations

    • Department of Radiology, University Hospital Gasthuisberg, University of Leuven, Leuven, Belgium.
  • ,
  • Frederik Maes, MSc, PhD

      Affiliations

    • Department of Electrical Engineering, University Hospital Gasthuisberg, University of Leuven, Leuven, Belgium
  • ,
  • Hein Heidbüchel, MD, PhD

      Affiliations

    • Department of Cardiology, University Hospital Gasthuisberg, University of Leuven, Leuven, Belgium

Received 24 January 2008; accepted 18 March 2008. published online 26 March 2008.

Background

We developed new methods for real time biplane integration of three-dimensional (3D) left atrial models with fluoroscopic images to assist in catheter ablation of atrial fibrillation (AF).

Objective

The purpose of this study was to quantitatively assess the accuracy of 3D fluoroscopy integration and to evaluate its clinical value when used as a single navigation tool for AF ablation.

Methods

Sixty patients underwent AF ablation under biplane fluoroscopic guidance after selective angiography of the four pulmonary veins. Computed tomography [CT]-based 3D models were integrated in the fluoroscopic framework using visual matching and landmark-based registration approaches. Integration accuracy was quantitatively assessed according to registration approach and different CT acquisition parameters (electrocardiogram [ECG] gating, respiratory phase). In 30 of the 60 patients (3D+ group), the integrated 3D model was used for real time 3D-augmented fluoroscopic catheter navigation, and the effects on procedural parameters and patient radiation dose were evaluated.

Results

Landmark-based registration resulted in superior 3D fluoroscopy integration accuracy compared with the visual matching approach (P <.001 for alignment error and alignment score). The effects of ECG gating and respiratory phase during CT acquisition on integration accuracy were small and clinically irrelevant. The use of 3D-augmented fluoroscopy in the 3D+ group was gauged as extremely helpful by the operator. It resulted in a significant reduction of fluoroscopy time (61 ± 18 minutes vs. 77 ± 26 minutes; P = .009) and a trend toward shorter procedure duration (230 ± 67 minutes vs. 257 ± 58 minutes; P = .06) versus conventional procedures. The systematic use of nongated cardiac CT in the 3D+ group resulted in an important reduction in total effective patient radiation dose due to CT+fluoroscopy (4 + 14 = 18 ± 8 mSv vs.17 + 16 = 33 ± 13 mSv; P <.001).

Conclusions

Biplane 3D-augmented fluoroscopy can be used as a safe and accurate stand-alone method to guide AF ablation procedures. The use of nongated cardiac CT substantially reduces total patient radiation dose without a relevant reduction in integration accuracy.

Keywords: Atrial fibrillation, Ablation, Image integration, Fluoroscopy, Computed tomography

To access this article, please choose from the options below

Login to an existing account or Register a new account.

  • Purchase this article for 31.50 USD (You must login/register to purchase this article)

    Online access for 24 hours. The PDF version can be downloaded as your permanent record.

  • Subscribe to this title

    Get unlimited online access to this article and all other articles in this title 24/7 for one year.

  • Claim access now

    For current subscribers with Society Membership or Account Number.

  • Visit SciVerse ScienceDirect to see if you have access via your institution.
 