Journal Home
Search for

Volume 7, Issue 2, Pages 193-198 (February 2010)


View previous. 12 of 33 View next.

System survival of nontransvenous implantable cardioverter-defibrillators compared to transvenous implantable cardioverter-defibrillators in pediatric and congenital heart disease patients

Andrew E. Radbill, MDa, John K. Triedman, MD, FHRSa, Charles I. Berul, MD, FHRSa, Francis Fynn-Thompson, MD, Joseph Atallah, MD, FRCPCa, Mark E. Alexander, MDa, Edward P. Walsh, MDa, Frank Cecchin, MDaCorresponding Author Informationemail address

Received 17 June 2009; accepted 11 October 2009. published online 16 October 2009.

Background

Nontransvenous (NTV) implantable cardioverter-defibrillator (ICD) systems with pericardial and/or subcutaneous coils are used in select pediatric and congenital heart disease patients who are not candidates for transvenous ICD leads. Outcomes with these hybrid configurations are not well understood.

Objective

The purpose of this study was to compare survival of NTV ICD systems to standard transvenous (TV) ICD systems.

Methods

We conducted a retrospective single-center study in which the TV group was matched to the NTV group 2:1 by type of cardiac disease and implant date.

Results

There were 39 patients in the NTV group and 78 matched in the TV group. Compared to the TV group, the NTV group was younger (median 7 vs 20 years) with a smaller body surface area at implant (0.9 vs 1.8 m2; both P <.001). Median follow-up in the NTV group was 17 months. System survival at 12, 24, and 36 months was 73%, 55%, and 49% in the NTV group versus 91%, 83%, and 76% in the TV group (P = .003). A multivariable Cox proportional hazards model including group, body surface area, and age at implant revealed the NTV group to be an independent predictor of system failure (hazard ratio 2.9, P = .04). Rate of total unanticipated interventions in the NTV group was 18 versus 6 per 1,000 person-months in the TV group. In patients with NTV systems, 23% received appropriate shocks and 18% received inappropriate shocks.

Conclusion

Survival of ICD systems using NTV defibrillation coils is significantly shorter than with TV ICD systems. Although NTV systems provide protection for this unique subset of patients, more durable options are needed.

a Department of Cardiology, Children's Hospital Boston, Harvard Medical School, Boston, Massachusetts

 Department of Cardiac Surgery, Children's Hospital Boston, Harvard Medical School, Boston, Massachusetts

Corresponding Author InformationAddress reprint requests and correspondence: Dr. Frank Cecchin, Department of Cardiology, Children's Hospital Boston, 300 Longwood Avenue, Boston, Massachusetts 02115

 Supported in part by the John R. Grey IV Cardiology Fellowship Fund at Children's Hospital, Department of Cardiology. Dr. Cecchin has received consulting fees and speaker honoraria from St. Jude Medical and Boston Scientific. Dr. Triedman has received consulting fees from Biosense Webster and Boston Scientific. Dr. Walsh has received speaker honoraria from St. Jude Medical and Boston Scientific.

PII: S1547-5271(09)01178-3

doi:10.1016/j.hrthm.2009.10.014


View previous. 12 of 33 View next.